21 CFR Part 11 (“Part 11”) is the US FDA regulation on electronic records and electronic signatures. It applies to companies shipping drugs or medical devices to the US, to those handling US clinical trial data, and indirectly to the vendors who supply them with systems. This introductory guide covers Part 11’s scope, the structure of the rule, the main requirements for electronic records and signatures, the “narrow interpretation” set out in the 2003 guidance, later developments, how it compares with Japan’s ER/ES guideline, and practical checkpoints for selecting and running systems.
What you will learn
- Which records Part 11 applies to (and how predicate rules fit in)
- The structure of Subparts A–C and the §11.10 controls (a)–(k)
- Electronic signature requirements, including certification to FDA
- What the 2003 “Scope and Application” guidance means
- Later developments, including the 2024 clinical investigations Q&A
- How Part 11 compares with Japan’s ER/ES guideline, and practical checkpoints
1. What Part 11 is
Part 11’s formal title is “Electronic Records; Electronic Signatures.” It is Part 11 of Title 21 of the Code of Federal Regulations, published in March 1997 and effective in August 1997. Within FDA regulations that assumed paper records and handwritten signatures, it sets the conditions under which electronic records and signatures can be treated as trustworthy equivalents of paper.
Part 11 does not itself say what must be recorded. That comes from individual regulations — drug GMP (21 CFR Parts 210/211), the device quality regulation (Part 820), GLP (Part 58), clinical trial rules and others — known as “predicate rules.” Part 11 defines what controls are needed when records required by predicate rules are created, maintained or submitted electronically.
2. Scope
| Where Part 11 typically applies | Where it typically does not |
|---|---|
| Records required by predicate rules that are kept electronically and relied on to perform regulated activities | Records created electronically but where printed paper is the official record and electronic data is not relied on (judged by actual practice) |
| Signatures required by predicate rules that are executed electronically | Records outside predicate rules (internal reference material, etc.) |
| Records submitted to FDA electronically | Paper records simply scanned and sent (depending on use) |
Note that “the printout is the official record, so Part 11 doesn’t apply” does not always hold. If decisions are made using electronic data, the electronic record will be treated as the true record in practice. The starting point is to define, system by system, which records are handled electronically.
3. Structure of the rule
| Subpart | Section | Content |
|---|---|---|
| A — General | §11.1 | Scope |
| §11.2 | Implementation | |
| §11.3 | Definitions (closed system, open system, electronic signature, etc.) | |
| B — Electronic records | §11.10 | Controls for closed systems |
| §11.30 | Controls for open systems (additional measures such as encryption) | |
| §11.50 | Signature manifestations (name, date/time, meaning) | |
| §11.70 | Signature/record linking | |
| C — Electronic signatures | §11.100 | General requirements (uniqueness, certification to FDA) |
| §11.200 | Electronic signature components and controls | |
| §11.300 | Controls for identification codes/passwords |
A closed system is one where access is controlled by the people responsible for the content of the records — most internal quality systems and ERP fall into this category. Open systems, which exchange data over the internet with outside parties, require additional measures under §11.30.
4. Controls for electronic records (§11.10)
| Item | Requirement | Practical example |
|---|---|---|
| (a) | Validation of systems | Verify accuracy, reliability, consistent performance and the ability to discern invalid or altered records |
| (b) | Ability to generate accurate and complete copies | Records can be provided in human-readable and electronic form during inspection |
| (c) | Protection of records throughout retention | Backup, archiving and restore procedures |
| (d) | Limiting system access | Only authorized individuals can use the system |
| (e) | Audit trails | Secure, computer-generated, time-stamped audit trails; changes must not obscure previous entries; retained as long as the records |
| (f) | Operational system checks | Enforce permitted sequencing of steps |
| (g) | Authority checks | Only authorized individuals can sign, alter records or perform operations |
| (h) | Device checks | Where appropriate, check the validity of the source of data input |
| (i) | Training | People who develop, maintain or use systems have the necessary education, training and experience |
| (j) | Written accountability for electronic signatures | Policies holding individuals accountable for actions under their electronic signatures |
| (k) | Control over systems documentation | Control of distribution, access and revision of operation and maintenance documentation |
5. Electronic signature requirements
- Manifestation (§11.50): signed records show the signer’s printed name, the date and time, and the meaning of the signature (review, approval, responsibility, etc.)
- Linking (§11.70): signatures are linked to their records so they cannot be excised, copied or transferred to falsify another record
- Uniqueness (§11.100): each electronic signature is unique to one individual and not reused or reassigned; identity is verified before assignment
- Certification to FDA (§11.100): the organization certifies to FDA in writing that its electronic signatures are intended to be the legally binding equivalent of handwritten signatures
- Components (§11.200): non-biometric signatures use at least two distinct components, such as an ID and a password
- ID/password controls (§11.300): uniqueness, periodic checks, deactivation of lost credentials, detection of unauthorized use
6. The 2003 guidance and the “narrow interpretation”
After Part 11 took effect, industry widely interpreted it as covering all electronic records, and compliance costs ballooned. In August 2003 FDA issued “Part 11, Electronic Records; Electronic Signatures — Scope and Application” to clarify its thinking.
- Part 11 is to be interpreted narrowly, applying to records required by predicate rules
- FDA exercises enforcement discretion for: validation, audit trails, record retention, record copying and legacy systems
- This does not mean these controls are unnecessary; they should be applied based on predicate rule requirements and a justified risk assessment
- Access controls, electronic signatures, operational and authority checks remain enforced
What it means in practice: since 2003, Part 11 compliance has shifted from “uniform controls on every record” to “controls proportionate to record criticality and risk.” That thinking carries through into GAMP 5’s risk-based approach and modern data integrity expectations.
7. Later developments
| When | Development | Significance |
|---|---|---|
| December 2018 | FDA “Data Integrity and Compliance With Drug CGMP” Q&A guidance | Sets out CGMP expectations on audit trail review, data deletion and more. See our data integrity guide |
| October 2024 | “Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations: Questions and Answers” finalized | Clarifies electronic systems, records and signatures in clinical trials in Q&A form; replaces the 2007 guidance for clinical investigations. Also clarifies that Part 11 compliance is not expected of electronic health records and other systems that serve only as sources of real-world data |
| September 2025 | “Computer Software Assurance (CSA)” guidance finalized (retitled in February 2026) | Scales validation effort to risk for software used in medical device production and quality systems. Aimed at devices, but influential on validation thinking in pharma as well |
The text of Part 11 has not changed since 1997, but how it is applied keeps evolving through guidance. Always read the related guidance alongside the rule.
8. Comparison with Japan’s ER/ES guideline
In 2005 Japan’s Ministry of Health, Labour and Welfare issued a notice on the use of electronic records and electronic signatures in applications for approvals and licenses (commonly called the ER/ES guideline). Like Part 11, it sets conditions for using electronic records and signatures.
| 21 CFR Part 11 | Japan ER/ES guideline | |
|---|---|---|
| Nature | Federal regulation (legally binding) | Ministry notice |
| Electronic records | §11.10 controls (a)–(k) | Organized around authenticity, readability and preservation |
| Electronic signatures | Manifestation, linking, uniqueness, certification to FDA | Identity verification, signature/record linking; no notification to the authority |
For systems used in both the US and Japan, an efficient approach is to design against the more detailed Part 11 requirements and then check for gaps from the ER/ES perspectives of authenticity, readability and preservation.
9. Practical checkpoints
When selecting a system
- Can audit trails be configured so they cannot be turned off? Are old values and reasons for change retained?
- Do electronic signatures show name, date/time and meaning on the record?
- Can permissions be granularly separated by role?
- Can records be output in both human-readable and electronic form?
When implementing
- Define for each system which records are electronic and where electronic signatures are used
- Translate Part 11 requirements into user requirements (URS) and verify them in validation (see our CSV guide)
- If electronic signatures are used, confirm certification to FDA has been submitted
When operating
- Proceduralize ID/password issuance, periodic review and deactivation for leavers
- Review audit trails at a frequency proportionate to risk
- Test backup and restore regularly
10. Common misconceptions
- “Buy a Part 11-compliant system and you’re done”: compliance cannot come from system features alone. Configuration, procedures, training and operation all matter — it is the user who complies
- “Every electronic record is in scope”: scope is records required by predicate rules, narrowly interpreted since 2003
- “Print it and sign it, and Part 11 doesn’t apply”: if decisions rely on the electronic data, the electronic record is in scope in practice
- “Audit trails are under enforcement discretion, so they’re optional”: from a data integrity standpoint, audit trails are effectively required
Summary
Part 11 sets the conditions for handling predicate-rule records and signatures electronically. Its core is the §11.10 controls and the electronic signature requirements, and since 2003 it has been applied with controls proportionate to record criticality and risk. For European computerised system requirements, see our EU GMP Annex 11 guide; for the thinking underlying both, see our data integrity guide.
Talk to us about Part 11 and system selection